Pressure Drop (cm WC)Flow through Sample (cm3/min)Face Velocity (cm/min)Challenge (particles/cm3)Output (particles/cm3)Corrected for Flow (particles/cm3)% Penetration
With hole
13.142100012.7323,02940.9440.940.178
10.077669.7624,59430.740.070.163
8.0746147.8225,25723.337.930.150
5.0223824.8729,26713.6935.830.122
3.0012282.9130,8507.1231.180.101
2.0261541.9633,7004.6430.100.089
1.07982.11.0534,6242.2527.400.079
0.54241.20.5333,3520.7919.160.057
0.22617.20.2234,4930.1810.470.030
Hole plugged
11.964100013.7024,34623.4823.480.0964
8.0086699.1727,51612.9919.410.0705
5.0034185.7330,3575.2212.480.0411
3.0352543.4831,1301.445.680.0182
2.0531722.3532,9080.553.210.0097
1.02185.31.1733,3420.0500.590.0018

Table I. Test data for coated Tyvek 1073B with 25-µm hole. Bold text indicates data taken directly from an instrument.

Return to the Article