
Originally Published EMDM November/December
2004
Industry News
ISO 14155 Clarifies Clinical Trial ProtocolsErik Swain
European clinical trials are now subject to more standardized procedures thanks
to the passage of ISO 14155: Clinical Investigation of Medical Devices for Human
Subjects. Anyone conducting a trial in Europe should now follow that document
instead of EN 540: Clinical Investigation of Medical Devices for Humans, experts
told attendees at Septembers MEDTEC Ireland conference in Galway.
In particular, the division of responsibilities between the investigators and
sponsors has been clarified, and there is more guidance on document and data
control, patient tracking, and informed consent procedures in ISO 14155.
There is more of an emphasis on quality assurance and quality control
for clinical investigations, said Danielle Giroud, president of d-Target
(Yverdon, Switzerland). The need to perform a detailed risk analysis
is emphasized.
She noted that the ISO document outlines all the elements that should be on
the informed consent forms, but added that, in terms of confidentiality
needs, you should go farther than what ISO says. In most cases, you just have
to explain what data [are] being collected and where [that data] will go. But
some countries, including Sweden, France, Norway, and Denmark, have additional
procedures. The Data Protection Directive says that what you do in one country
is OK for all the others. That is not always true.
The ISO document demands careful accounting for all patients who withdraw from
a clinical study. You must document them very clearly, Giroud said.
Otherwise, authorities might ask if you are trying to hide something.
Among the elements that must be included in a clinical investigation plan (CIP)
are:
The objective of the study and the supporting rationale.
The design of the clinical investigation: what is the product being compared
to, and why?
Statistical considerations expected to be taken into account.
Methods for reporting adverse events experienced in the trial.
Conditions for an early termination of the investigation.
Its important to the competent authorities that all this information
be presented clearly, said Barbara Tucker, medical assessor for medical devices
at the Irish Medicines Board (Dublin).I feel that [a plan] should read
like a story, she said. Often leaps are made from one section to
another without justification of scientific rationale.
She also emphasized that the product design should be frozen once the CIP has
been submitted to the competent authority. There is no time for the competent
authority to revisit amendments, she said. Avoid this by getting
to a design freeze.
Further refinements of the standard are needed, however, Giroud said. It
just says that investigators shall allow auditing. But by whom? The sponsor?
The competent authority? Another issue, she said, is that there is no
consensus on how ethics committees should be constituted.
Another goal is to harmonize ISO 14155 better with the International Council
on Harmonizations good clinical practice documents. However, Tucker said,
the documents include a lot of pharmaceutical terminology.
Copyright ©2004
European Medical Device Manufacturer


